SCIO S ETUP
VF8 DELTA/DELTA XL/KAPPA 23-13
To access the Scio etCO
2
* setup menu
Click on the etCO
2
* parameter box
or
1. Press the
Menu fixed key.
2. Click on
Patient Setup.
3. Click on
Parameters to display a list of available parameters.
4. Click on
etCO2*.
Quick Reference Table – etCO2* Setup
Click on the following items to execute etCO
2
* setup functions:
NOTE: The monitor does not alarm for etCO
2
* or inspiratory and expiratory agent
limit violations until it has established a valid respiratory rate.
The etCO
2
*
Setup Menu
Menu Item Description Settings
Scale Sets the etCO
2
* waveform scale. • 40, 60, 80 mmHg
Respiratory
Sweep Speed
Sets waveform sweep speed on screen display. • 6.25, 12.5, 25, 50 mm/s
Agent Display Displays a separate agent parameter box;. Ghosted
if the monitor is displaying combined MultiGas
parameter box (see page 23-22).
•ON
• OFF
Pressure
Comp.
Sets the compensation for ambient atmospheric
pressure.
•Auto
• 760 mmHg
RRc* Apnea
Time
Sets the time that the monitor waits before reporting
a cessation of breathing as an apnea event.
• OFF, 10, 15, 20, 25, 30 s
Apnea
Archive
Allows you to store and/or record an alarm event
automatically for apnea. You can later review stored
alarms on the Event Recall screen.
• OFF, Record, Store
(default), Str./Rec.
MultiGas Zero Manually zeroes the Scio module (parameter values
disappear during zeroing).
Not applicable