25-10 DELTA/DELTA XL/KAPPA VF8
25 B
ISPECTRAL INDEX (BISX) MONITORING
BIS: Sensor expired The sensor is expired. Remove the expired sensor and replace
with a new/compatible sensor.
BIS: Check Sensor /
Impedance
Sensor fault
Defective sensor cable
Make sure the sensor is properly
applied to the patient, then perform a
manual impedance check to continue
monitoring. If this fails, do one of the
following:
• Disconnect and examine the sensor.
• Replace the sensor, if necessary.
• Replace the sensor cable.
• Call your hospital’s technical
personnel or DrägerService.
BIS: Pod disconnected The BISx pod is not
connected to the monitor.
Check the cables and connections.
Replace the cables, if necessary.
BIS: Pod Failure BISx pod self-test failed
BISx pod hardware failure
BISx pod software failure
Defective sensor
Check the BISx pod connection.
Replace the sensor, then disconnect
and reconnect the pod to continue
monitoring.
Call your hospital’s technical
personnel or DrägerService.
Duplicate devices
connected
Multiple BISx pods or MIB BIS
is connected.
Disconnect the unwanted duplicate pod
or device.
BIS: SRS Final use SRS cable last use in
progress.
Replace the SRS cable after current
use.
BIS: Replace Sensor Unknown sensor type
Too many uses/sensor worn
out
Invalid sensor
Defective sensor cable
SRS no more uses available
(When SRS is connected
after last use.)
Remove the unknown/worn out/invalid
sensor and replace with a new/
compatible sensor.
Replace the sensor cable.
Replace the SRS cable before
continuing.