25-10 DELTA/DELTA XL/KAPPA VF8
25 B
ISPECTRAL INDEX (BISX) MONITORING
BIS: Sensor expired The sensor is expired. Remove the expired sensor and replace
with a new/compatible sensor.
BIS: Check Sensor /
Impedance
• Sensor fault
• Defective sensor cable
• Make sure the sensor is properly
applied to the patient, then perform a
manual impedance check to continue
monitoring. If this fails, do one of the
following:
• Disconnect and examine the sensor.
• Replace the sensor, if necessary.
• Replace the sensor cable.
• Call your hospital’s technical
personnel or DrägerService.
BIS: Pod disconnected The BISx pod is not
connected to the monitor.
• Check the cables and connections.
• Replace the cables, if necessary.
BIS: Pod Failure • BISx pod self-test failed
• BISx pod hardware failure
• BISx pod software failure
• Defective sensor
• Check the BISx pod connection.
• Replace the sensor, then disconnect
and reconnect the pod to continue
monitoring.
• Call your hospital’s technical
personnel or DrägerService.
Duplicate devices
connected
Multiple BISx pods or MIB BIS
is connected.
Disconnect the unwanted duplicate pod
or device.
BIS: SRS Final use SRS cable last use in
progress.
Replace the SRS cable after current
use.
BIS: Replace Sensor • Unknown sensor type
• Too many uses/sensor worn
out
• Invalid sensor
• Defective sensor cable
• SRS no more uses available
(When SRS is connected
after last use.)
• Remove the unknown/worn out/invalid
sensor and replace with a new/
compatible sensor.
• Replace the sensor cable.
• Replace the SRS cable before
continuing.
Message Possible Cause Suggested Action