PICCO SETUP INCLUDING IBP
VF8 DELTA/DELTA XL/KAPPA 26-9
3. Insert an arterial thermodilution catheter into a suitable artery (femoral,
brachial or axillary artery) of the patient. Take care that all air is evacuated
from the blood pressure lumen.
4. Connect the pressure line of the monitoring kit to the lumen of the
thermodilution catheter.
1
1
CVP catheter1
6
CO port
2
Injectate temperature sensor
housing1
7
Pulsion blood pressure transducer1
3
p-CO thermistor cable
8
Delta, Delta XL, or Kappa PodCom
connection (Delta XL shown)
4
p-CO intermediate cable
9
10-pin cable to transducer
5
PiCCO pod
10
Pulsion arterial thermodilution catheters1
1
Catheters and blood pressure transducers required for use with the PiCCO pod are only
available from Pulsion Medical Systems. Contact your local sales representative for