26-20 DELTA/DELTA XL/KAPPA VF8
26 P
ULSE CONTOUR CARDIAC OUTPUT (PICCO) MONITORING
Available functions, present on Thermodilution setup menu, are described in the
following table.
p-BT <value>
p-IT <value>
Read only display of p-BT and p-IT
values.
NOTE: If p-BT or p-IT values are out
of range, the values will display “***”.
p-BT range: –3 to –31°C (27 to 88°F)
p-IT range: 25 to 43°C (77 to 109°F)
Not applicable
PBW <value>
PBSA <value>
Read only display of PBW and PBSA
values.
Not applicable
PiCCO Alarms Allows user to set PiCCO parameter
alarms via alarm limits menu. See
page 5-6.
NOTE: PiCCO alarms annunciate/
flash in the parameter box even if
alarming parameter is not selected to
be displayed in the parameter box.
Not applicable
NOTE: The user can perform a TD measurement even if ART pressure is not
monitored.
Thermodilution Setup Menu
Menu Item Description Available Settings
Show All
Parameters
Displays all the PiCCO (continuous
and discrete) parameters in a
separate window
Not applicable
Parameters Setup Allows you to designate which
parameter displays as Parameter 1,
Parameter 2 and Parameter 3.
Parameter 1:
GEDVI (default),
p-CO, p-CI, GEDV, ITBV, ITBVI,
EVLW, EVLWI, CFI, PVPI, GEF
Parameter 2:
EVLWI (default),
p-CO, p-CI, GEDV, GEDVI, ITBV,
ITBVI, EVLW, CFI , PVPI, GEF
Parameter 3
GEF (default),
p-CO, p-CI, GEDV, GEDVI, ITBV,
ITBVI, EVLW, EVLWI, CFI, PVPI
Pulse Contour Parameters Setup Menu
Menu Item Description Available Settings