EasyManua.ls Logo

Dräger Infinity Delta - Page 465

Dräger Infinity Delta
614 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 26-27
ART/CVP: Did
Not Zero
Transducer zeroing failed
because of:
Excessive signal noise
A non-static waveform
Keep all tubing motionless, then rezero.
Change the transducer.
Check the stopcock, then rezero.
ART/CVP: Did
Not Zero - Offset
Error
Transducer zeroing failed
because the static pressure
was too high or too low.
Rezero the transducer.
Loosen and retighten the transducer dome,
then rezero the transducer.
Replace the transducer and rezero.
PiCCO: Unknown
Catheter
Unknown/unsupported catheter
type is being used.
Unplug the unknown catheter and replace
with the correct catheter type (see
“Precautions” on page 26-6.)
PiCCO: ART
Catheter
Unplugged
The arterial catheter is
unplugged.
Injectate probe thermistor
open circuit.
Replug the catheter/cable.
NOTE: If the ART catheter is disconnected then
re-connected, TD values are restored and pulse
contour values are updated.
PiCCO: Cal
Required
NOTE: This
message will
continue to
display until a
successful
PiCCO calibration
with a valid ART
pressure value is
completed.
Thermodilutions were
performed, then the ART
pressure transducer was
connected.
Check the transducer and connections.
Reconnect and recalibrate, if necessary, with
a valid ART pressure.
Replace the transducer, if necessary.
ART/CVP
Transducer
disconnected
The ART/CVP pressure
transducer has been
disconnected.
Check if necessary transducer and
connections.
Reconnect, if necessary, and rezero.
Replace the transducer, if necessary.
P-CO Check
Injectate Probe
The injectate probe is
disconnected.
Injectate probe thermistor
open circuit/malfunction.
The p-IT value is out of range.
Make sure the probe is connected properly
and repeat the measurement.
Check the cable and/or the injectate
temperature sensor housing.
Make sure the injectate temperature housing
is not exposed to extraneous warming source
(for example warming blanket).
Change the sensor cable and/or injectate
temperature housing, if necessary.
PiCCO Injectate
probe Failure
The injectate temperature
sensor cable is loose or
defective.
The injectate temperature
sensor housing is defective.
Make sure the probe is connected properly
and repeat the measurement.
Check the cable and/or injectate temperature
sensor housing.
Change the sensor cable and/or injectate
temperature housing, if necessary.
Message Possible Cause Suggested Action

Table of Contents

Other manuals for Dräger Infinity Delta

Related product manuals