28-8 DELTA/DELTA XL/KAPPA VF8
28 P
ERIPHERAL DEVICES AND ASSOCIATED SOFTWARE
Supported Ventilator Parameters
The following table lists the parameters for all ventilators except Evita Infinity V500
whose parameters are listed separately on page 28-9.
NOTE: For information on calculating parameter values, see the documentation for
the peripheral device you are using
CAUTION: The parameter abbreviations on the bedside monitor may differ from those
displayed on the ventilator (refer to the table on page 28-9).
The bedside monitor displays pressure units in cmH
2
O while the ventilator displays
pressure units in mbar (1 cmH
2
O being equal to 1 mbar approximately).
Parameter Bedside
monitor
label
Bedside and
Infinity
network
units of
measure
Where displayed
Peak Airway Pressure PIP cmH
2
O•Show All Parameters screen
• Paw (Flow and
Volume) parameter box
• Vent view (Infinity
CentralStation)
Mean Airway Pressure MAP cmH
2
O
Positive End Expiratory
Pressure
PEEP cmH
2
O
Minute Volume (Expired) MVe L/min • Show All Parameters screen
• Vent (Flow and
Volume) parameter box
• Vent view (Infinity
CentralStation)
Respiration Rate (Breathing
Frequency)
RRv bpm
Tidal volume TVe l
Plateau (Pause) Pressure
NOTE: Pause appears blank on
the monitor for Fabius GS.
PAUSE cmH
2
O•MIB Show All
Parameters screen
• Vent view (Infinity
CentralStation)
Inspired:Expired Ratio I:E Not
applicable
Inspired Oxygen iO
2
%
Minute Volume (Inspired) MVi L/min
Inspired Tidal Volume TVi l
Inspiration Time InspT% sec
Spontaneous Minute
Ventilation
1
MVe s 1/min
Spontaneous Respiration Rate
(Spontaneous Breathing
Frequency)
1
RRs breaths/min