28-32 DELTA/DELTA XL/KAPPA VF8
28 P
ERIPHERAL DEVICES AND ASSOCIATED SOFTWARE
. The Warmer2 menu has the following selections.
Babytherm Status Messages
Menu Item Description Available Settings
Show All
Parameters
Displays all parameters
# Parameter
Boxes
Select number of Parameter Boxes 1, 2 (default)
NOTE: Warmer2 Parameter box is
not displayed when the number of
Parameter Boxes is set to 1
Parameter 1 Selects up to 3 parameters for display in
warmer parameter box
T1skin, T2skin, Tmatt, HtrPwr
(default), None
Parameter 2 T1skin, T2skin, Tmatt, HtrPwr,
None (default)
Parameter 3 T1skin, T2skin,Tmatt, HtrPwr,
None (default)
Warmer
Mini Trend
Allows you to select parameters for
display in mini trends (See page 6-7 for
more information on mini trend display).
T1skin, T2skin, Tmatt, HtrPwr
(default)
Message Condition Suggested Action
Duplicate Device
Connected
Multiple devices are connected. Make sure only one incubator is
connected to the monitor.
T1skin Out of Range The T1skin value is out of range
(high or low).
Check the patient and treat, if
necessary.
Check the connections.
Unplug and reconnect the
incubator.
Power-cycle the monitor or
undock and redock the monitor.
Call the manufacturer.
T2skin Out of Range The T2skin value is out of range
(high or low).
Tmatt Out of Range The Tmatt value is out of range (high
or low).
HtrPwr Out of Range The HtrPwr value is out of range
(high).