29-8 DELTA/DELTA XL/KAPPA VF8
29 C
LEANING AND DISINFECTING
Airway Adapter
1. Rinse airway adapters in a warm soapy solution, then soak them in a liquid
disinfectant.
2. Dry with a lint-free cloth, making sure adapter windows are dry and free of
any residue before they are used.
Sidestream Sampling Pump (etCO
2
only)
The etCO
2
module and pod contain a small pump that draws air from the nasal cannula,
through the sidestream airway adapter, and out the exhaust port. Suggested cleaning proce-
dures are outlined below.
The following fluids are acceptable for cleaning:
Isopropyl alcohol.
A 5.25 % water solution (by weight) of sodium hypochlorite (bleach).
The following items are required for cleaning the etCO
2
module or pod:
A 60 cc catheter-tip syringe.
A 2-foot section of 1/8- or 3/16-inch tubing to drain off fluid after it passes
through the etCO
2
pump.
A receptacle to receive the fluid after it drains.
CAUTION:
To reduce the risk of infection, remember that airway and flow sensors are for
single patient use only and cannot be sterilized. To avoid blocking the sampling
tube, do not attempt to sterilize the pod by immersing it in liquids.
Do not attempt to sterilize the etCO
2
/Respiratory Mechanics pod by immersing
it in liquids.
WARNING: Used sampling tubing, T-connectors, and water
traps could be contaminated and must be handled and
disposed of with care. Infection hazard may be present.
Dispose of these items in accordance with local regulations.