MONITORING A CCESSORIES
VF8 DELTA/DELTA XL/KAPPA B-31
Masimo SET
®
– SpO2 pod
Physical Attributes
Size (H x W x T) 39 x 58 x 135 mm (1.5 x 2.3 x 5.3 in)
Weight 0.15 kg (0.33 lbs)
Connections X8:for power and connection to monitor
Cable connection: for intermediate cable to Masimo sensors
Power source: Powered directly from monitor
Mode of operation: Continuous
Degree of protection
against electric shock:
Type CF
Protection against
ingress of water:
IPX1 (per IEC 60529)
Environmental Requirements
Temperature range: Operating: 5 to 45 °C (41to 113 °F)
Storage: –20 to 60 °C (–4 to 140 °F)
Relative humidity: Operating:10 to 95 %, non-condensing
Storage: 10 to 95 %, (with packaging)
Atmospheric pressure Operating:485 to 795 mmHg (64.7 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)