MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-47
Alarm limit resolution: etCO
2
, iCO
2
: 0.1 mmHg (0.01 kPa)
RRc: 1 breath/min
Apnea detection? Module:
Adult and pediatric:No
Neonatal:Yes
Pod: Yes in all three patient categories
RRc range (pod): Mainstream:0-149 breaths/min
Sidestream:0-69 breaths/min
Accuracy:±1 breath/min
Rise time: Mainstream:<100 ms
Sidestream:<200 ms
Delay time: Mainstream:<100 ms
Sidestream:<450 ms
Total system response time: Rise time plus delay time
Cross-sensitivity Compensation Error: Gas measurement accuracy was tested at the gas
levels listed in the following table. The additional worst case error when compensation is correctly
selected (for barometric pressure, O2, and N2O) for the actual fractional gas constituents that are
present is:
0 to 40 mmHg: ± 2 mmHg additional error
41 to 70 mmHg: ± 3 % additional error
71 to 99 mmHg: ± 5 % additional error
Additional notes regarding cross-sensitivity compensation errors:
Xenon and Helium: Constantly bias etCO2 values >= 38 mmHg by up to –5 mmHg
Desflurane: At concentrations > 5%, constantly biases etCO2 values >= 38 mmHg by up to
+4 mmHg
Quantitative Effects Measured At
Gas or Vapor Gas Level (Gas levels in % volume fraction)
Nitrous oxide 60
Halothane 4
Enflurane 5
Isoflurane 5
Sevoflurane 5
Xenon 80
Helium 50
Metered dose inhaler
propellants
Not specified for use
Desflurane 15
Ethanol 0.1
Isopropanol 0.1