MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-51
Flow/Volume Parameters (Derived)
Measurement range,
Adult/Pediatric
Raw i, Raw e
Cdyn
C20/Cdyn
RSBI
I:E
Ti, Te
InspT%
WOBimp
0 to 100tocmH2O/L/s
0 to 500tomL/cmH2O
Not applicable
0 to 1000tobpm/L
6:1, 1:6
0 to 15tosec
0 to 100
0 to 8 J/L/s
Measurement range,
Neonatal
Raw i, Raw e
Cdyn
C20/Cdyn
RSBI
I:E
Ti, Te
InspT%
WOBimp
0 to 500 cmH2O/L/s
0 to 500 mL/cmH2O
0 to 1
Not applicable
6:1, 1:6
0 to 15 sec
0 to 100
Not applicable
Flow/Volume/CO2 Parameters (Derived)
Measurement range
Adult
VCO2
TVCO2
TValv s, TValv m, TValv
MValv s, MValv m,MValv
PeCO2
TVd aw
TVd/TVaw
0 to 400 mL
0 to 120 mL
0, 100 to 3000 mL
0 to 60 L/min
0 to 50 mmHg
0 to 500 mL
0 to 0.75
Measurement range
Pediatric
VCO2
TVCO2
TValv s, TValv m, TValv
MValv s, MValv m,MValv
PeCO2
TVd aw
TVd/TVaw
0 to 400 mL
0 to 120 mL
0, 10 to 400 mL
0 to 30 L/min
0 to 50 mmHg
0 to 500 mL
0 to 0.75
Measurement range
Neonatal
VCO2
TVCO2
TValv s, TValv m, TValv
MValv s, MValv m,MValv
PeCO2
TVd aw
TVd/TVaw
0 to 400 mL
0 to 120 cmL
0, 2 to 250 mL
0 to 8 L/min
0 to 50 mmHg
0 to 250 mL
Not applicable
Airway Adapter and sensor dimensions
Type Deadspace Inner diameter Weight without tubing
Flow Neonatal <1 mL 5mm <7 g
Flow Adult/Pediatric 6.5 mL 15 mm <9 g
CO2/Flow Neonatal <1 mL 5mm <11 g
CO2/Flow Pediatric <4 mL 10 mm <11 g