2-8 DELTA/DELTA XL/KAPPA VF8
2 M
ONITOR SETUP
Monitor Options submenu
Date & Time Sets the date and time displayed in the
lower right potion of the main screen
NOTES:
• An internal battery powers the monitor’s
clock even when the monitor is turned
off.
• This option is not available when the
monitor is connected to network, since
network date and time are set at the
central station.
• Changing the time does not affect other
time-related functions such as timers
and time stamps.
To set the monitor date and time:
1) Click on Date & Time.
2) Click on Current Date. A data entry
screen appears.
3) Click on Day, scroll to the correct
date, and click.
4) Repeat Step 3 for Month and Year.
5) Click on Accept to confirm or on
Cancel to return to submenu.
6) Click on Current Time to set the
time, using the same method
described in steps 3 and 5.
Speaker
Volumes
Allows you to set the volume for alarms,
pulse tones, and attention tones
• Click on Alarm Volume to set the
volume of alarms (10-100 % in
increments of 10).
• Click on Pulse Tone Volume to set
the volume of the pulse tone (OFF-
100 % in increments of 10 after 5).
• Click on Attention Tone Volume to
set the volume of attention tones
(OFF-100 % in increments of 10 after
5).
Trend Setup Allows you to configure trend display. Submenu (see chapter 6, for detailed
information).
Recordings Allows you to configure and assign
recorders.
Submenu (see chapter 7, for
information).
Biomed Provides access to technical and clinical
logs and service menus.
Submenu (see page 2-21).
Unit Manager Allows the unit manager, physician or
head nurse to configure monitoring
functions for the clinical staff.
Submenu (see page 2-9).
OR Configures monitor to meet the special
needs of the operating room environment.
Submenu (see page 2-12).
The Main Menu
Menu Item Description Available Settings