MONITORING S PECIFICATIONS
VF8 DELTA/DELTA XL/KAPPA B-57
BIS Parameters
Via BISx Pod
Temperature
BIS index: Parameter Label:BIS
Measurement Range:0 to 100
Display Resolution:1
Signal quality index: Parameter Label:SQI
Measurement Range:0 to 100%
Display Resolution:1
Suppression ratio: Parameter Label:BSR
Measurement Range:0 to 100%
Display Resolution:1%
EMG: Parameter Label:EMG
Measurement Range:0 to 100 dB (in the frequency range from 70-110
Hz)
Display Resolution:1 dB
Total power: Parameter Label:PWR
Measurement Range:40 to 100 dB (frequency range 0.5-30 Hz)
Display Resolution:1 dB
Spectral edge
frequency:
Parameter Label:SEF
Measurement Range:0.5 to 30 Hz
Display Resolution:0.1 Hz
Burst Count: Parameter Label:BCT
Measurement Range:0 to 30
Display Resolution:1
Parameter display: Absolute temperature, delta temperature (with hemodynamic pods)
Measurement range: Absolute:–5 to 50 °C (23 to 122 °F)
Delta: 0 to 55 °C (32 to 131 °F)
Resolution: 0.1 °C
Accuracy: Absolute:±0.1 °C
Delta: ±0.2 °C
Average response time: <2.5 seconds
Probe accuracy: ±0.1 °C 0 to 50 °C
Degree of protection
against electric shock:
Type CF
Defibrillation protection: Defibrillation-proof applied part per IEC 60601-1
NOTE: Range and accuracy values are also applicable to hemodynamic pods.