SPECIALTY MENUS
VF8 DELTA/DELTA XL/KAPPA 2-17
Audio Alarm
Reminder
Allows you to set a reminder when the
Alarm Volume feature is set to OFF.
ON (default) – When the Alarm
Volume feature is set to OFF, a
reminder tone sounds every
30 seconds at 50 % volume.
NOTE: For monitors in OR mode:
At the end of an alarm silence or
an alarm off period, if the alarm
condition is still active, the
parameter box will flash and an
appropriate reminder tone (high,
medium, or low) sounds every 30
seconds at 50 % volume.
OFF – There is no reminder
tone when Alarm Volume is set
to OFF.
The Code Setup submenu
This menu allows the unit manager to configure the monitor for quick emergency response. Open
the Unit Manager menu (page 2-14), click on Code Setup, then select and execute functions as
described in this table. For more information, see page 5-6.
Continuous
Record
NOTE: If no recorder is available, the
recording request remains pending for
later printing.
Yes (default) – Provides a
continuous recording when you
press the Code fixed key.
No – No continuous recording is
started when you press the
Code fixed key.
Continuous NBP NOTE: You must attach the NBP cuff and
display the NBP parameter box before
requesting NBP measurements.
Yes – Starts continuous NBP
measurements when you press
the Code fixed key.
No (default) – No NBP
measurements are started when
you press the Code fixed key.
Alarm Volume
OFF
Allows you to lower the alarm volume to
its minimum setting (see page 2-8 and
page 2-13) when you press the Code
fixed key.
Yes – Allows you to lower the
alarm volume.
No (default) – The alarm volume
is not affected when you press
the Code fixed key.
The Unit Manager Menu
Menu Item Description Available Settings

Table of Contents