3-14 DELTA/DELTA XL/KAPPA VF8
3 N
ETWORK APPLICATIONS
Quick Reference – Remote View Setup
Menu Item Description Settings
Select
Remote Bed
Displays up to two waveforms and
parameter boxes of a remote bed.
If the remote bed is not in alarm,
the top two waveforms are
displayed on the local bed,
otherwise the alarming waveform
occupies the bottom channel.
NOTES:
The monitor continuously
updates the remote bed label,
patient name, and alarm or status
messages.
The remote display appears on
the bottom half of the screen so
that you can continue to display
the local monitor’s top
waveform(s), parameter box(es),
and message area.
1) Press the Fast Access fixed key.
2) Click on Remote View.
3) Select Remote Bed to display a list of
all beds in the monitoring unit.
4) Click on the label of the bed you wish to
view.
5) Press the Main Screen fixed key to
return to Main Screen, or click on Select
Remote Bed to return to the Remote View
menu.
NOTE: To access Remote View functions,
see the section “Remote View Screen” on
page 3-15.
Alarm Group Assigns the monitor an alarm
group number (0 - 255), allowing
you to restrict the number of
messages received over the
network from the central station or
other bedsides
1) Press the Fast Access fixed key.
2) Click on Remote View.
3) Click on Alarm Group.
4) Click on the number of the desired
alarm group.
Auto Dual View Configures the monitor to display
any remote bed in alarm that is part
of the local bed’s alarm group
1) Press the Fast Access fixed key.
2) Click on Remote View.
3) Click on Auto Dual View.
4) Click to toggle ON or OFF.
NOTE: Beds in the same alarm group
continue to post messages in the alarm
group’s network message area when Auto
Dual View is disabled. If you do not want
to display messages at a particular bed,
place the bed in its own alarm group by
selecting an unassigned Alarm Group
number.

Table of Contents

Other manuals for Dräger Infinity Delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Delta and is the answer not in the manual?

Dräger Infinity Delta Specifications

General IconGeneral
Display Size10.4 inch
Resolution800 x 600 pixels
Battery TypeLithium-ion
Parameters MonitoredECG, SpO2, NIBP, Invasive Blood Pressure, Temperature, CO2, Cardiac Output
ECGYes
SpO2Yes
NIBPYes
Invasive Blood PressureYes
TemperatureYes
CO2Yes
Cardiac OutputYes
Power Supply100 to 240 V AC, 50/60 Hz
ConnectivityWLAN
Operating Temperature0 to 40 °C
Storage Temperature-20 to 60 °C
Humidity15 to 95 % relative humidity, non-condensing
AlarmsAudible and visual alarms