VF8 DELTA/DELTA XL/KAPPA 3-17
Steps: Activating Privacy mode
1. Press the Menu fixed key.
2. Click on
Privacy.
3. Press the Main Screen fixed key to return to the main screen.
WARNING: When a bedside monitor is in privacy mode,
audible alarms only sound at the Infinity CentralStation (the
bedside monitor does not provide audible alarms or activate
its alarm bar).
NOTE: Alarm bar not applicable to SC 7000/8000/9000XL or Kappa.