4.3.5 Set the overlap in Route
Percent Overlap sets how much each new average overlaps the previous average when
taking a measurement. The higher the overlap percentage, the less newly acquired data is
needed to create a spectrum. A higher percentage decreases the data collection time.
The default overlap, 67 percent, is acceptable for most situations.
Procedure
1.
Activate a route.
2. From the Route Data Collection screen, press ALT > F1 User Setup > F6 Percent Overlap.
3. Enter a percent between 0 and 99.
The default is 67 percent.
4. Press Enter.
4.3.6 Set the plot type for live route data
Select Live Display sets the type of plot to display on the Route Data Collection screen during
data collection.
Procedure
1. Activate a route.
2. From the Route Data Collection screen, press ALT > F1 User Setup > F8 Select Live
Display.
3. Use the up and down arrow keys to select an option.
Option Description
Status Display the overall and average count. This option has the fastest acquisition. The
default is Status.
Spectrum Display a spectrum for each point.
Waveform Display a waveform for each point.
Dual Displays a waveform and spectrum for each point.
4. Press Enter.
4.3.7 Set the overall mode
Overall mode includes additional frequencies in your measurement points.
You can set the Overall mode in AMS Machinery Manager. If the setting for Overall is
different in the analyzer, the setting defined in AMS Machinery Manager is used.
Procedure
1.
Activate a route.
Route