10.6.3 Set the weight plane radius
You can set the trial and trim weight placement radius. Weight plane radius ensures you
use the same radius in the job. Use this option as a reference and for future setups only.
This setting is optional. The analyzer does not use weight plane radius to calculate the
balance correction.
Procedure
1.
From the Balance main menu, press F3 Job Setup.
2. Press F5 Weight Plane Setup or Enter until the Weight Plane Setup screen appears.
3. Ensure the correct weight plane displays at the top of the screen.
4. Press F6 Enter Weight Radius.
Enter a value between 0.00 and 650 inches for the plane. The default is 0
(Unknown).
5. Enter the radius, and press Enter.
6. Repeat these steps for each weight plane.
10.7 Acquire Data
After you set up your job, you can acquire the runout (optional), reference run, and trial
run data.
The Balance program has a recommended sequence, but you can acquire data on any
measurement run. Acquire Data consists of several screens. On the first screen, select the
measurement. After you complete a measurement, an X appears in the brackets next to
the measurement.
Balance
322 MHM-97432 Rev 7

Table of Contents

Other manuals for Emerson CSI 2140

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson CSI 2140 and is the answer not in the manual?

Emerson CSI 2140 Specifications

General IconGeneral
BrandEmerson
ModelCSI 2140
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals