EasyManua.ls Logo

Fluke 990DSL - Page 89

Fluke 990DSL
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
POTS Testing and Fault Location
The POTS Toolbox
4
4-25
Single-Sided Fault Location Test Results
To locate a single-sided fault, do the following:
1. Connect the strap as shown in Figure 4-12, then press K.
2. If the tester was turned off during the strap connection process:
Turn it on, and select RFL again from the POTS Toolbox.
Select A(
Continue Prev. Test), then press K.
If the strap is not detected, a “Strap Connection Error” message is
displayed.
If the strap is properly connected, the tester performs a “fault null”
operation and displays a screen similar to the one in Figure 4-14.
acy23s.eps
Figure 4-14. Two-Wire RFL Test Results: Tip Ground
Figure 4-14 shows the conductor resistance values to the fault, and from
the fault to the strap (shown in the diagram as a solid line).

Table of Contents

Related product manuals