990DSL
Users Guide
4-28
3. Enter the length of the first section (five digits maximum, with no
commas), using the same units of measure (feet or meters) selected for the
Language configuration, then press L.
4. PressG to move the cursor to the Section 1
Gauge field.
5. Press A(
Edit), press F or H for the desired gauge, then press L.
6. PressG to move the cursor to the
Load Coil field.
7. Press A(
Edit), and press F or H to select Y if a load coil terminates the
section or
N if it does not. Then, press L.
8. PressG to move the cursor to the
Strap Dist. field.
9. If the Section 1 length entered is the actual known distance to the strap,
press F or H to select
Y. Otherwise select N, and press L.
10. PressG to move the cursor to the Section 2
Length field.
11. Repeat Steps 2-7 until you enter information for all of the sections. When
you finish, press M to return to the
RFL results screen.
Note
It is important to make sure that the last section that you enter
contains the strap position. If an ā€œ
Insufficient Length Enteredā€
message is displayed when you press the
M
key to exit the Multi-
gauge Entry screen, it is because the sum of the entered lengths is
shorter than the Distance-to-Strap length measured by the tester. In
this situation, first verify that the entered
Length and Gauge
information is correct. If so, increase the length you entered for the
last section. This change does not affect the accuracy of the results
but allows agreement between the entries and measurements.