Troubleshooting
6-19
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0205524
Warning
Read IQ param
failed, the value
is over limit
Image processing parameter
over limit. Image quality may
be affected. Please restart the
system. If this issue persists,
please call for service.
check IQ config file
0205525
Warning
Write IQ
parameters
failed
Image processing parameter
save failure. Image quality may
be affected. Please restart the
system. If this issue persists,
please call for service.
check IQ config file
0201501
Warning
Start callback
failure
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
Start capture failed.
1. Gige camera may be
abnormal
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
0201502
Warning
End callback
failure
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. Gige camera may be
abnormal
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
install software.
0201503
Warning
Init camera
buffer failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
End capture failed.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-