Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-20
ID
Warning
Description
(Displayed in User Interface)
failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
between camera
and workstation.
2. Check ETH1 port on
workstation mainboard.
3. Replace the camera.
configuration
file failed
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
between camera
and workstation.
2. Check ETH1 port on
workstation mainboard
thread failed
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
thread priority
failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-