1  Introduction Important information on wall thickness measurement using the DMS Go+
1-8 Edition 4 (05/2014) DMS Go+
1.3 Important information on wall 
thickness measurement 
 
using the DMS Go+
Please always read the following information 
about the special features in wall thickness 
measurement using the DMS Go+. It is abso
-
lutely necessary that you observe this infor-
mation in order to always obtain correct 
measurement results. 
Probe zero offset
High differences in temperature 
If the differences in temperature between the storage 
and inspection site are high, it is necessary that you wait 
approx. 2 minutes before you use the instrument after 
connecting the probe.
Temperatures below –10 °C
The probe zero offset does not always function correctly 
at temperatures below –10 °C. You should therefore 
carry out a 2-point calibration and repeat it in the case 
of major temperature jumps.
Couplant residues
In order to ensure a correct probe zero offset, make sure 
that you always remove the couplant residues before 
carrying out a new measurement after decoupling the 
probe.
Measuring accuracy 
Keep in mind that the measuring accuracy is not identi-
cal with the display accuracy.
The measuring accuracy depends on the following fac-
tors:
● Temperature
● Probe delay line
● Constancy of sound velocity
● Uniformity of surface quality