7 Special functions of the DMS Go+ TC Measurement mode Auto-V
7-6 Edition 4 (05/2014) DMS Go+
7.2 Measurement mode Auto-V
Applications
The Auto-V function is used for the simultaneous deter-
mination of sound velocity and wall thickness on non-
coated test objects.
The simultaneous determination of sound velocity is not
possible in wall thickness measurements on coated test
objects.
ATTENTION
A prerequisite for correct measurement re-
sults is a good homogeneity of the material in
the immediate vicinity of the measurement
location.
As opposed to that, large-area inhomogene-
ities have an effect on the calculation of wall
thickness.
Choosing the probe
To be able to use the Auto-V function, you have to con-
nect a special 4-element probe to the DMS Go+.
The following probes from GE are suitable for Auto-V:
TC-560
Activating Auto-V
Connect a suitable probe to the instrument.
In the function group PROBE&CAL, select the func-
tion PROBE.
Choose the connected probe. The instrument is im-
mediately adjusted.
Select the function TG MODE and choose the option
AUTO-V.