5  Operation Other functions
5-52 Edition 4 (05/2014) DMS Go+
– In the function group PROBE&CAL, select the func-
tion TEMP COMP. 
– Press one of the left or right arrow keys in order to 
choose the option ENABLE.
– Select the function CAL TEMP and set the tempera-
ture of the calibration block.
– Select the function CUR TEMP and set the tempera-
ture of the test object.
– Select the function TEMP COEF and set the tem-
perature coefficient.
– Finally, press the function key 1 (function HOME) in 
order to exit the function. The settings become effec
-
tive immediately.
The DMS Go+ can display both the uncom-
pensated and the compensated thickness 
value at the same time. To achieve this, con
-
figure the data boxes accordingly (see Sec-
tion Configuring data boxes, page 5-13).
Display freeze (FREEZE)
You can freeze the display screen of the DMS Go+. The 
A-scan and the displayed thickness value are then fro
-
zen and no longer updated. 
This function is available in all measurement modes and 
display views.
When the function is active, a corresponding icon is dis-
played at the top left corner of the display screen (see 
Section Status display icons, page 0-5).
– In the function group MEASURE, select the function 
FREEZE.
– Press one of the left or right arrow keys in order to 
switch the function on.
You can assign the function FREEZE to the 
function key 4 (
see Section Assignment of 
function key 4, page 5-7).