5 Operation Other functions
5-52 Edition 4 (05/2014) DMS Go+
In the function group PROBE&CAL, select the func-
tion TEMP COMP.
Press one of the left or right arrow keys in order to
choose the option ENABLE.
Select the function CAL TEMP and set the tempera-
ture of the calibration block.
Select the function CUR TEMP and set the tempera-
ture of the test object.
Select the function TEMP COEF and set the tem-
perature coefficient.
Finally, press the function key 1 (function HOME) in
order to exit the function. The settings become effec
-
tive immediately.
Note
The DMS Go+ can display both the uncom-
pensated and the compensated thickness
value at the same time. To achieve this, con
-
figure the data boxes accordingly (see Sec-
tion Configuring data boxes, page 5-13).
Display freeze (FREEZE)
You can freeze the display screen of the DMS Go+. The
A-scan and the displayed thickness value are then fro
-
zen and no longer updated.
This function is available in all measurement modes and
display views.
When the function is active, a corresponding icon is dis-
played at the top left corner of the display screen (see
Section Status display icons, page 0-5).
In the function group MEASURE, select the function
FREEZE.
Press one of the left or right arrow keys in order to
switch the function on.
Note
You can assign the function FREEZE to the
function key 4 (
see Section Assignment of
function key 4, page 5-7).