5  Operation Other functions
5-50 Edition 4 (05/2014) DMS Go+
Calibration reminder
Certain measurement tasks require a regular calibration 
of the gauge. 
The reminder function of the DMS Go+ uses three crite-
ria from which you can choose one:
● Time in minutes
● Number of readings
● Power On and Off
If you switch on the reminder functions of DMS Go+, a 
reminder icon is displayed at the top left corner of the 
display screen once the selected criterion is reached 
(
see Section Status display icons, page 0-5).
– In the function group PROBE&CAL, select the func-
tion CAL REMINDER. 
– Press one of the left or right arrow keys. A selection 
menu is displayed.
– Press one of the left or right arrow keys in order to 
select the required criterion.
– If necessary, press one of the up or down arrow keys 
in order to define additional settings for the selected 
criterion.
– Press one of the left or right arrow keys in order to 
set the additional data.
– Finally, press the function key 2 (function CONFIRM) 
in order to save the data. The selection menu disap
-
pears, and the selected settings are displayed at the 
function CAL REMINDER.