5 Operation Other functions
5-50 Edition 4 (05/2014) DMS Go+
Calibration reminder
Certain measurement tasks require a regular calibration
of the gauge.
The reminder function of the DMS Go+ uses three crite-
ria from which you can choose one:
Time in minutes
Number of readings
Power On and Off
If you switch on the reminder functions of DMS Go+, a
reminder icon is displayed at the top left corner of the
display screen once the selected criterion is reached
(
see Section Status display icons, page 0-5).
In the function group PROBE&CAL, select the func-
tion CAL REMINDER.
Press one of the left or right arrow keys. A selection
menu is displayed.
Press one of the left or right arrow keys in order to
select the required criterion.
If necessary, press one of the up or down arrow keys
in order to define additional settings for the selected
criterion.
Press one of the left or right arrow keys in order to
set the additional data.
Finally, press the function key 2 (function CONFIRM)
in order to save the data. The selection menu disap
-
pears, and the selected settings are displayed at the
function CAL REMINDER.