92
ENGLISH
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situaons. Electromagnec propagaon is aected by absorpon and
reecon from structures, objects and people.
a: Field strengths from xed transmiers, such as base staons for radio (cellular / cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theorecally with
accuracy. To assess the electromagnec environment due to xed RF transmiers, and electromagnec site survey
should be considered. If the measured eld strength in the locaon in which the Paent Monitor is used exceeds the
applicable RF compliance level above, The Paent Monitor should be observed to verify normal operaon. If abnormal
performance is observed, addional measures may be necessary, such as re-orienng or relocang the Paent
Monitor.
b: Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3V/m.
Table 4
Frequency Range and Level: RF wireless communicaon equipment
TTest Frequency (MHz) Modulaon
Minimum immunity
Level (V/m)
immunity Level
Applied (V/m)
385 **Pulse Modulaon: 18 Hz 27 27
450
*FM + 5 Hz deviaon: 1 kHz sine
**Pulse Modulaon: 18 Hz
28 28
710
745
780
**Pulse Modulaon: 217 Hz 9 9
810
870
930
**Pulse Modulaon: 18 Hz 28 28
1720
1845
1970
**Pulse Modulaon: 217 Hz 28 28
2450 **Pulse Modulaon: 217 Hz 28 28
5240
5500
5785
**Pulse Modulaon: 217 Hz 9 9