 Vigorous movement of the paent, strong ambient light, or extreme electro-surgical interference may also aect
the SpO
2
measuring accuracy.
 DO NOT stare at the light of SpO
2
sensor (infrared is invisible) when switch it on, for the infrared may do harm to
the eye.
 The informaon, such as the range of the peak wavelengths and maximum opcal output power of the light by
the SpO
2
sensor can be especially useful to clinicians.
 Always observe the plethysmogram (waveform), which is auto-scaled (normalized). When the measured signal
is inadequate, the waveform will be not smooth or irregular, the SpO
2
reading may be unlikely true or displayed
with “--”, and even a technical alarm will be generated. If in doubt, rely on your clinical judgement, rather than
the monitor readout.
 Please do not use the SpO
2
sensor and the monitor when doing the MRI imaging, or burn may be caused by
faradism.
 For disposal SpO
2
sensor, If the sterile packaging is damaged, do not use it any more.
 Check the SpO
2
sensor and cable before use. Do NOT use the damaged SpO
2
sensor.
 Before each use, surface-clean sensor and cable with a so gauze pad by saturang it with a soluon such as 70%
isopropyl Ethanol. If low-level disinfecon is required, use a 1:10 bleach soluon.
 When the temperature of SpO
2
sensor is abnormal, do not use it any more.
 Please do not allow the cable to be twisted or bent.
 Please do not use nail polisher or other cosmec product on the nail.
 The ngernail should be of normal length.
 The SpO
2
sensor cannot be immersed into water, liquor or cleanser completely, because the sensor has no
capability to resist the harmful ingress liquid.
 Do not disinfect any SpO
2
sensor by irradiaon, steaming, or ethylene oxide.
 Carefully route cables to reduce the possibility of paent entanglement or strangulaon.
The clinical study for SpO
2
measurement accuracy was done on human subjects according to Standard ISO 80601-
2-61.
A funconal tester or SpO
2
simulator can not be used to assess the accuracy of the oximeter or a SpO
2
sensor.
However, it can be used to check how accurately a parcular oximeter is reproducing the given calibraon curve.
Before tesng the oximeter by a funconal tester, please rstly ask the manufacturer which calibraon curve is
used, if necessary, request the manufacturer for its dedicated calibraon curve and download it into the tester.
10.3 Apply the Sensor
1.Select an appropriate sensor and probe according to the module type and paent category..
2. Apply the sensor to the proper site of the paent.
3. Select an appropriate adapter cable according to the connector type and plug this cable into the SpO
2
connector.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gima PC-3000 and is the answer not in the manual?

Gima PC-3000 Specifications

General IconGeneral
BrandGima
ModelPC-3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals