Gima PC-3000 - Operation Instructions; Screen Display for Arrhythmia Event List

Gima PC-3000
282 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
ENGLISH
4.5 Screen Display for Arrhythmia Event List
Figure 4.13 Arrhythmia Event List Screen
4.5.1 Operaon Instrucons
The structure is similar to the recall screen. We will cover each funcon key below.
Start: This buon is used to start and end the system Arrhythmia detecon. The default is OFF. When the Arrhythmia is
not ON, the “Learn” key is disabled. Press this key and the monitor enters learning stage before detecng arrhythmia.
The “Startchanges to “End”. Press it again to end the learning. When “Learn” key change to yellow from gray, it indicates
the learning has nished. When the ARR detecon begins aer learning, the monitor will automacally detect the ARR
event. If an ARR event is detected, the ECG waveform with corresponding ARR event will be displayed in the 3
rd
trace of
waveform, shown in Figure 4.13.
When the monitor get Reset or the paent ID has been changed, the ARR detecon needs to be re-learned.
Learn: Because the Arrhythmia detecon is based on the paern of normal ECG waveform, which is built up during
learning stage. When the paent changes, the ARR detecon could be incorrect, so re-learning is needed. To get beer
result of the learning funcon, it is recommended to start learning while a segment of good quality ECG waveform
appears during the ECG monitoring.
View: Press this key, and the selected record in the Arrhythmia event list, which is to be reviewed, becomes the green.
Rotate the “Navigaon Knob” to choose the record and display the corresponding waveform in the 3
rd
trace of waveform.
Press the knob again to exit.
Exit: press this key to exit the Arrhythmia detecon screen and return to the System Menu screen.