32
ENGLISH
4.5 Screen Display for Arrhythmia Event List
Figure 4.13 Arrhythmia Event List Screen
4.5.1 Operaon Instrucons
The structure is similar to the recall screen. We will cover each funcon key below.
Start: This buon is used to start and end the system Arrhythmia detecon. The default is OFF. When the Arrhythmia is
not ON, the “Learn” key is disabled. Press this key and the monitor enters learning stage before detecng arrhythmia.
The “Start” changes to “End”. Press it again to end the learning. When “Learn” key change to yellow from gray, it indicates
the learning has nished. When the ARR detecon begins aer learning, the monitor will automacally detect the ARR
event. If an ARR event is detected, the ECG waveform with corresponding ARR event will be displayed in the 3
rd
trace of
waveform, shown in Figure 4.13.
When the monitor get Reset or the paent ID has been changed, the ARR detecon needs to be re-learned.
Learn: Because the Arrhythmia detecon is based on the paern of normal ECG waveform, which is built up during
learning stage. When the paent changes, the ARR detecon could be incorrect, so re-learning is needed. To get beer
result of the learning funcon, it is recommended to start learning while a segment of good quality ECG waveform
appears during the ECG monitoring.
View: Press this key, and the selected record in the Arrhythmia event list, which is to be reviewed, becomes the green.
Rotate the “Navigaon Knob” to choose the record and display the corresponding waveform in the 3
rd
trace of waveform.
Press the knob again to exit.
Exit: press this key to exit the Arrhythmia detecon screen and return to the System Menu screen.