17.9 Other Technical Specicaons
1. Power supply: AC100V-240V, 50/60Hz, 60VA; Internal power supply (Lithium baery): DC 14.8V
2. Mode of operaon: Connuous
3. Applied Part: ECG lead wire, SpO
2
sensor, TEMP sensor and Cu
4. Display mode: TFT color LCD
5. Alarming mode: audible & visible alarm
6. Communicaon: Net port
17.10 Classicaon
Safety standard: IEC 60601-1
The type of protecon against electric shock Class I and Internally powered equipment
The degree of protecon against electric shock Debrillaon-proof BF and Debrillaon-proof CF applied parts
Electro-Magnec Compability Group I, Class A
The protecon against harmful ingress of liquid
or parculate maer
IPX2
The device is not intended for use in an oxygen rich environment.
The surface of the device can be cleaned and disinfected with 75% Ethanol, no need to sterilize.
17.11 Operang Environment
1. Ambient temperature range: 5°C~40°C
Relave humidity: 15%~85%, non-condensing
Atmospheric pressure: 70kPa ~106.0kPa
Power Voltage: (100-240) VAC
Power frequency: 50Hz/60Hz
2. This equipment should be situated in a place protected against direct sunlight, so as to prevent overheang inside the
equipment.
3. The device should be stored and used within specied temperature, humidity and atmospheric pressure range, or it
may cause damage to the device or inaccurate measurement result.
4. If the device gets wet by accident, the operator should NOT power it on directly unl it has been air-dry enough to
avoid any damage to it.
5. Do not use this equipment in an environment with toxic or inammable gas.
6. This equipment should be placed on a stand or at plaorms, so as to prevent possible shock.
7. Do not use this equipment in combinaon with any equipment other than those expressly permied in the manual.
8. The monitor is debrillator discharge proof and can be used with electrosurgical unit. But when the device is used
together with debrillator or electrosurgical equipment, the user (doctor or nurse) should keep the paent under close
surveillance for his/her safety. Refer to the following funcon descripon for specic protecve measures or notes.
9. Make sure that the equipotenal grounding terminal is grounded correctly.
10. Do not use mobile phone nearby, so as to avoid strong radiant eld interference.