85
ENGLISH
17.9 Other Technical Specicaons
1. Power supply: AC100V-240V, 50/60Hz, 60VA; Internal power supply (Lithium baery): DC 14.8V
2. Mode of operaon: Connuous
3. Applied Part: ECG lead wire, SpO
2
sensor, TEMP sensor and Cu
4. Display mode: TFT color LCD
5. Alarming mode: audible & visible alarm
6. Communicaon: Net port
17.10 Classicaon
Safety standard: IEC 60601-1
The type of protecon against electric shock Class I and Internally powered equipment
The degree of protecon against electric shock Debrillaon-proof BF and Debrillaon-proof CF applied parts
Electro-Magnec Compability Group I, Class A
The protecon against harmful ingress of liquid
or parculate maer
IPX2
The device is not intended for use in an oxygen rich environment.
The surface of the device can be cleaned and disinfected with 75% Ethanol, no need to sterilize.
17.11 Operang Environment
1. Ambient temperature range: 5°C~40°C
Relave humidity: 15%~85%, non-condensing
Atmospheric pressure: 70kPa ~106.0kPa
Power Voltage: (100-240) VAC
Power frequency: 50Hz/60Hz
2. This equipment should be situated in a place protected against direct sunlight, so as to prevent overheang inside the
equipment.
3. The device should be stored and used within specied temperature, humidity and atmospheric pressure range, or it
may cause damage to the device or inaccurate measurement result.
4. If the device gets wet by accident, the operator should NOT power it on directly unl it has been air-dry enough to
avoid any damage to it.
5. Do not use this equipment in an environment with toxic or inammable gas.
6. This equipment should be placed on a stand or at plaorms, so as to prevent possible shock.
7. Do not use this equipment in combinaon with any equipment other than those expressly permied in the manual.
8. The monitor is debrillator discharge proof and can be used with electrosurgical unit. But when the device is used
together with debrillator or electrosurgical equipment, the user (doctor or nurse) should keep the paent under close
surveillance for his/her safety. Refer to the following funcon descripon for specic protecve measures or notes.
9. Make sure that the equipotenal grounding terminal is grounded correctly.
10. Do not use mobile phone nearby, so as to avoid strong radiant eld interference.