48
ENGLISH
counng-down me, otherwise, it displays the date. Move the cursor on this text display area can bring up the edit box
of “Alarm volume” seng.
6.3.6 Changing Alarm Volume
To set the alarm sound volume.
Step 1: Select “System Menu”→“Setup”→“SYS”→“Alarm Volume”.
 When the alarm sound is switched o, the monitor will give no audible alarm tones even if a new alarm occurs.
Therefore the user should be very careful about whether to switch o the alarm sound or not.
 Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low level
may result in a hazard to the paent. Always keep the paent under close surveillance.
 Pausing or switching o alarms may result in a hazard to the paent, please be very careful.
 If occurrence of multiple alarm signal at the same time, the monitor will only show the high priority alarm in the form of
audible alarm and alarm lamp. Meanwhile, all alarm signals information including message descriptions and numerical
highlight etc. will be shown respectively and simultaneously on the screen.
 It is suggested that the users should not change the alarm volume lower than the factory default setting if close and
constant attention could not be paid to the patient, otherwise the negligence of alarm event might cause irreversible harm
to the patient.
 During the alarm silence period, any new alarm event can activate the audible alarm again and the audible alarm function
resumes normal state.
 When all global sound and audio sound are disabled (the sound volume is zero), then the device will give out sound like
beep, beep, beep as the reminder signal. The sound goes o every hour.
6.4 Understanding the Alarm Seng
Select “Menu” and select parameter (such ECG, SpO
2
, NIBP, TEMP, etc.). you can review and set alarm limits, alarm
switches, high/low alarm limit for each parameter.
1. The volume of audible alarm is adjustable, the other properties of the alarm cannot be adjusted by the user. Such as alarm
priority setting, alarm lamp ashing and so on. In addition, all alarms in this patient monitor are “non-latched”, that is to say,
when the alarm event disappears, the corresponding alarm will automatically stop. The alarm volume range is shown as below:
 High: 0dB80dB (The distance between device front and test instrument is 1m)
 Medium: 0dB75dB (The distance between device front and test instrument is 1m)
 Low: 0dB70dB (The distance between device front and test instrument is 1m)
2. Alarm settings are non-volatile, that means the previous settings will still sustain even the patient monitor is powered o
including unexpected power failure and manual reboot.