44
ENGLISH
 Exit: return to the System Setup screen.
5.6 SpO
2
Related Sengs
On “System Menu” screen, select “Setup”“SpO
2
” to enter into SpO
2
related sengs.
Figure 5.6 SpO
2
related sengs
 SpO
2
Hi/Lo: High and Low alarm limits for SpO
2
 Pulse Hi/Lo: High and Low alarm limits for pulse rate.
 Exit: return to the System Setup screen.
5.7 Respiraon Related Sengs
On “System Menu” screen, select “Setup”“RESP” to enter into Respiraon
related sengs.
Figure 5.7 Respiraon related sengs
 Gain: Respiraon amplicaon/gain, 4 opons, x1/2, x1, x2, and x4. The default is x2
 Speed: Respiraon waveform sweeping speed, 2 opons 6.25mm/s and 12.5 mm/s. The default is 12.5 mm/s
 Apnea: meout seng for apnea alarm. Opons: “OFF” and any number from 5 to 120. The step is 1 and its unit is
second. When the breathing acon is not detected aer the me longer than this set period, the trace for respiraon
waveform will display warning “Apnea xxx second”. The default is “OFF”.