Chapter 15 Maintenance
In case of trouble of this machine in the service, follow the instrucons below to eliminate the problem rst. If the
aempt fails, refer to the dealer in your local area or the manufacturer. Refer to the detailed provisions in contract for the
warranty period of the main unit and the accessories of this monitor.
15.1 Daily Examinaon
Before using the monitor, the checks below should be carried out:
 Check the monitor for any mechanical damage;
 Inspect the exposed parts and the inserted parts of all the cables and the accessories;
 Examine all the funcons of the monitor that are likely to be used for paent monitoring, and ensure that it is
in good working condion;
 Make sure that the monitor is grounded properly.
 Pay close aenon to the uctuaon of the local power supply voltage. A power voltage regulator is
recommended when necessary.
In case any indicaon of damage about the funcon of the monitor is detected and proven, it is not allowed to apply it to
the paent for any monitoring. Please contact the local dealer or our company, and we are to oer the best soluon as
soon as possible for your sasfacon.
15.2 Roune Maintenance
At each rounely maintenance or the yearly maintenance, the monitor can be thoroughly inspected by qualied
personnel, including performance and safety examinaons. This monitor is designed with life cycle of 5 years. It is strongly
recommended to use the product which is sll within its life cycle, or it may cause inaccurate measurement. In order to
ensure its long service life, please pay aenon to the maintenance.
 If the hospital fails to carry out a sasfactory maintenance program about the monitor, it may get disabled and
harm the paent’s safety and health.
 Perform protecve grounding impedance, leakage current and insulaon resistance test.
 In case of ECG cable/lead wires damage or aging, please replace the cable or lead wires.
 If there is any indicaon of cable and transducer damage or they deteriorate, they are prohibited from any further
use.
 The Monitor is calibrated in the factory before sale, so there is no need to calibrate it during its life cycle. Any paent
simulators should not be used to validate the accuracy of blood pressure and oxygen saturaon measurement,
they can only be used as funconal testers to verify its precision.
 The accuracy of ECG signal amplicaon can be veried by the built-in 1mV calibraon signal.
 The accuracy of pressure measurement and air leakage in pneumac system can be veried by means of the built-
in pressure vericaon funcon and a precision pressure meter, please refer to the related chapter in Part 2 of
the user manual for detail operaon.
 The SpO
2
simulator can not be used to verify the SpO
2
measuring accuracy, which should be supported by the clinical
study conducted by inducing hypoxia on healthy, non-smoking, light-to-dark-skinned subjects in an independent
research laboratory. However it is necessary for the user to use SpO
2
simulator for roune vericaon of precision.
 Please note that the specic calibraon curve (so called R-curve) should be selected when use of SpO
2
simulator,e.g.
for Index 2 series SpO
2
simulator from Fluke Biomecidal Corporaon, please set “Make” to “DownLoadMake:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gima PC-3000 and is the answer not in the manual?

Gima PC-3000 Specifications

General IconGeneral
BrandGima
ModelPC-3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals