6.4.1 High and Low Alarm Setting Range
Parameter
Seng range
High limit Low limit
HRbpm (Low limit+1)~350 0~(High limit-1)
S-TmV (Low limit+0.01)~2.50 -2.5~(High limit-0.01)
SpO
2
% (Low limit+1)~100 0~(High limit-1)
PRbpm (Low limit+1)~300 0~(High limit-1)
RRrpm (Low limit+1)~150 0~(High limit-1)
TEMP1(℃) (Low limit+0.1)~60.0 0~(High limit-0.1)
TEMP2(℃) (Low limit+0.1)~60.0 0~(High limit-0.1)
TD(℃) 0.0~5.0
NIBP (Unit)
mmHg
Adult Pediatric Neonate
SYS
High limit Low limit+1)~280 Low limit+1)~200 Low limit+1)~135
Low limit 29~(High limit-1 29~(High limit-1 29~(High limit-1
MAP
High limit Low limit+1)~242 Low limit+1)~165 Low limit+1)~110
Low limit 20~(High limit-1 20~(High limit-1 20~(High limit-1
DIA
High limit Low limit+1)~232 Low limit+1)~150 Low limit+1)~100
Low limit 10~(High limit-1 10~(High limit-1 10~(High limit-1
NIBP (Unit)
mmHg
Adult Pediatric Neonate
SYS
High limit Low limit+0.1)~37.3 Low limit+0.1)~26.7 Low limit+0.1)~18.0
Low limit 3.9~(High limit-0.1 3.9~(High limit-1 3.9~(High limit-0.1
MAP
High limit Low limit+0.1)~32.3 Low limit+0.1)~22.0 Low limit+0.1)~14.7
Low limit 2.7~(High limit-0.1 2.7~(High limit-0.1 2.7~(High limit-0.1