23
ENGLISH
The environment where the equipment is used shall be reasonably free from noises, vibraon, dust, corrosive, ammable
and explosive substances. If the equipment is installed in a cabinet, sucient space in front and behind shall be le for
convenient operaon, maintenance and repair. Moreover, to maintain good venlaon, the equipment shall be at least 2
inches (5cm) away from around the cabinet.
When the equipment is moved from one place to another, condensaon may occur as a result of temperature or humidity
dierence. In this case, never start the system before the condensaon disappears.
3.2 Geng Started
3.2.1 Connecng to Power Source
1. Using AC Power Source
Make sure that the AC power supply is (100-240) VAC, 50Hz/60Hz.
Use the power cable provided by the manufacturer. Insert one end of it to the AC power input of the
monitor and the other end to the three-pin outlet of the power source with protected-earth.
To eliminate potenal dierences, the monitor has a separate connecon to the equipotenal grounding
system. Connect one end of the provided ground wire to equipotenal grounding terminal on the rear of
the monitor, and connect the other end to one point of the equipotenal grounding system.
Cauon: 1. Ensure that the monitor is grounded correctly.
2. If you have any doubt to the grounding layout and its performance, you must use the built-in baery to
power the monitor.
Aer the mains power supply has been interrupted while power switch remains in the “on” status and is restored
aer a period of me that is longer than 30 seconds, the monitor will run by the last sengs when restarng the
monitor.
The monitor is applicable to connect to the public mains power network.
2. Using Baery
The following steps should be followed to install the baery:
Step 1: open the baery cover;
Step 2: pull out the baery cable and connect it to the baery pack;
Step 3: push the baery pack into the baery compartment and lock it;
Step 4: close the baery cover.
Cauon: it’s beer to recharge the baery aer it is used up, and the charging me should be 12~15 hours long.
Baery life: Provided that a baery is new and fully charged, the minimal working me of the monitor with accessories
connected is declared in the table below:
Name Baery life
Paent Monitor More than 120min
NOTE: when the device is working, It takes at least 10 hours to charge baery from exhaust state to 90% charged.
The provided baery of the monitor must be recharged aer transportaon or storage. So if the monitor is