C - EMC Compliance
Essenal Performance
The monitor has the following essenal performance in an environment of electromagnec environment specied below:
Operang mode, accuracy, funcon, alarm
Warnings:
The instrument conforms to the requirements of IEC60601- 1 - 2 , EN 60601-1-2 standards for electromagnec
compability.
The user shall install and use the EMC informaon provided in the random le.
Portable and mobile RF communicaon equipment may aect the performance of the instrument, avoid strong
electromagnec interference when using, such as close to the mobile phone, microwave oven, etc .
The guidance and manufacturer's declaraon are detailed in the table below .
The instrument should not be close to or stacked with other equipment. If it must to be close to or stacked, it should
be observed and veried to be able to operate normally under its conguraon.
In addion to the cables sold by the instrument manufacturer as spare parts for internal components, the use of other
accessories and cables may result in increased emission or reduced immunity.
Table 1
Guidance and manufacturers declaraon-electromagnec emission
The Paent Monitor is intended for use in the electromagnec environment specied below. The customer or the
user of the Paent Monitor should assure that it is used in such an environment.
Emissions test Compliance
Electromagnec environment-
guidance
Conducted emissions
CISPR 11
Group 1
Class A
The Paent Monitor uses RF energy
only for its internal funcon.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
Radiated emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Class A
The Paent Monitor suitable for
use in all establishments, including
domesc establishments and those
directly network that supplies
buildings used for domesc
purposes.
Voltage uctuaons/icker emissions
IEC61000-3-3
Complies