90
ENGLISH
Table 2
Guidance and manufacturer’s declaraon-electromagnec emission
The Paent Monitor is intended for use in the electromagnec environment specied below. The customer or the user
of the Paent Monitor should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level
Electromagnec environment
-guidance
Electrostac
discharge (ESD)
IEC61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ± 15 kV
air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ± 15 kV
air
Floors should be wood, concrete or
ceramic le. If oors are covered
with synthec material, the relave
humidity should be at least 30%
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
Supply lines
±1kV for Input a.c. Power
Ports
±2kV for power
Supply lines
±1kV for Input a.c. Power
Ports
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±0.5 kV, 1kV line (s) to line(s)
±0.5 kV, ± 1 kV, ±2kV line(s)
to earth
±0.5 kV, 1kV line (s) to line(s)
±0.5 kV, ± 1 kV, ±2kV line(s)
to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interrupons and
voltage variaons on
power supply input
lines
IEC61000-4-11
<5% UT (>95% dip in UT) for
0.5 cycle
<40% UT (60% dip in UT) for
5 cycles
<70% UT (30% dip in UT) for
25 cycles
<5% UT (>95% dip in UT) for
5 s
<5% UT (>95% dip in UT) for
0.5 cycle
<40% UT (60% dip in UT) for
5 cycles
<70% UT (30% dip in UT) for
25 cycles
<5% UT (>95% dip in UT) for
5 s
Mains power quality should be
that of a typical commercial or
hospital Environment .If the user
of the paent monitor requires
connued operaon during
power mains interrupons, it is
recommended that the paent
monitor should be powered from
an uninterrupble power supply or
a baery
Power frequency
(50Hz/60Hz) magnec
eld
IEC61000-4-8
30A/m 30A/m
Power frequency magnec
elds should be at levels
characterisc of a typical locaon
in a typical commercial or hospital
environment.