Table 2
Guidance and manufacturers declaraon-electromagnec emission
The Paent Monitor is intended for use in the electromagnec environment specied below. The customer or the user
of the Paent Monitor should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level
Electromagnec environment
-guidance
Electrostac
discharge (ESD)
IEC61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ± 15 kV
air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ± 15 kV
air
Floors should be wood, concrete or
ceramic le. If oors are covered
with synthec material, the relave
humidity should be at least 30%
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
Supply lines
±1kV for Input a.c. Power
Ports
±2kV for power
Supply lines
±1kV for Input a.c. Power
Ports
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±0.5 kV, 1kV line (s) to line(s)
±0.5 kV, ± 1 kV, ±2kV line(s)
to earth
±0.5 kV, 1kV line (s) to line(s)
±0.5 kV, ± 1 kV, ±2kV line(s)
to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interrupons and
voltage variaons on
power supply input
lines
IEC61000-4-11
<5% UT (>95% dip in UT) for
0.5 cycle
<40% UT (60% dip in UT) for
5 cycles
<70% UT (30% dip in UT) for
25 cycles
<5% UT (>95% dip in UT) for
5 s
<5% UT (>95% dip in UT) for
0.5 cycle
<40% UT (60% dip in UT) for
5 cycles
<70% UT (30% dip in UT) for
25 cycles
<5% UT (>95% dip in UT) for
5 s
Mains power quality should be
that of a typical commercial or
hospital Environment .If the user
of the paent monitor requires
connued operaon during
power mains interrupons, it is
recommended that the paent
monitor should be powered from
an uninterrupble power supply or
a baery
Power frequency
(50Hz/60Hz) magnec
eld
IEC61000-4-8
30A/m 30A/m
Power frequency magnec
elds should be at levels
characterisc of a typical locaon
in a typical commercial or hospital
environment.