83
ENGLISH
3. RESP rate alarm limit seng range: High: 1rpm~150rpm; Low: 0rpm~149rpm.
4. Alarm tolerance: ±1rpm
17.3 TEMP
1. TEMP measuring range: 21.0 ~50.0
2. TEMP measuring accuracy: ±0.2
for the range from 25 to 45 , 0.4 for other ranges.
3. TEMP responding me: ≤150s
4. TEMP minimum measuring me: ≥130s
5. Measuring site: body surface
6. Mode of operaon: direct mode
7. Unit: and °F
8. TEMP alarm limit seng range: High: 0 ~60 ; Low: 0 ~59.9 .
9. Tolerance: ±0.1
17.4 NIBP
1. Measuring method: Oscillometric Technique
2. Pneumac pressure measuring range: 0 mmHg~300mmHg
3. Accuracy of pressure measurement: ±3 mmHg
4. Typical measurement me: <30 seconds (adult cu)
5. Measurement me on the average: < 90 seconds
6. Air release me while the measurement is canceled: <2 seconds (typical adult cu)
7. Inial cu inaon pressure
Adult: <150 mmHg; Infant: <120 mmHg; Neonate: <70 mmHg Tolerance: ±5 mmHg
8. Overpressure protecon limit
Adult: 300 mmHg; Infant: 240mmHg; Neonate: 150 mmHg
9. NIBP measurement range:
Pressure (unit) Adult Infant
Neonate
SYS
mmHg 40~275 40~200 40~135
kPa 5.336.7 5.326.7 5.318.0
MAP
mmHg 20~230 20~165 20~110
kPa 2.730 2.722.0 2.714.6
DIA
mmHg 10~210 10~150 10~95
kPa 1.328 1.320.0 1.312.7
10. NIBP accuracy:
Maximum mean dierence: ±5 mmHg
Maximum Standard deviaon: 8 mmHg
11. Measurement mode: Manual, Auto, STAT
12. NIBP Alarm seng range: see Secon Alarms
17.5 SpO
2
1. Transducer: dual-wavelength LED
Wavelength: Red light: 660 nm, Infrared light: 905 nm