17.12 Storage
If the equipment will not be used for long period of me, wipe it clean and keep it in the packaging, which shall be kept
in a dry and good venlaon place free from dust and corrosive gases
Storage environment: ambient temperature: -20~60°C
relave humidity: 10%~95%
atmospheric pressure: 53kPa~106kPa
17.13 Transportaon
This monitor should be transported by land (vehicle or railway) or air in accordance with the contractual terms. Do not
hit or drop it with force.
Transportaon environment: ambient temperature: -20~60°C
relave humidity: 10%~95%
atmospheric pressure: 53kPa~106kPa
17.14 Packaging
The product is packed in high quality corrugated cartons with foam inside to protect the apparatus against damage in the
handling process.
Gross weight: Details see the indicaon on the outer package
Dimension: Details see the indicaon on the outer package
Chapter 18 Troubleshoong
Note: In case of trouble of this machine in service, follow the instrucons below to eliminate the problem rst. If the
aempt fails, contact the dealer in your local area or the manufacturer.
Do NOT open the monitor cabinet without permission
18.1 No Display on the Screen
Shut down the machine and unplug the power cable. Use a universal meter to check if the outlet has proper voltage,
if the power cable is in good condion, and if the power cable is properly connected with this apparatus or outlet.
Remove the fuse from the back cover of this machine, and make sure it is in good condion. If all of above is in good
condion, the display screen may be malfuncon.
18.2 Excessive ECG Signal Interference or too Thick Baseline
1. Check if the plate electrodes are properly located, and if valid plate electrodes are used.
2. Check whether the lead wires are properly inserted. If no ECG curve displayed, check if the ECG lead wires are broken.
3. Make sure the mains outlet has standard grounding wire.
4. Check if the grounding wire of the apparatus properly grounded.
18.3 No Blood Pressure and Pulse Oxygen Measures
1. Check if the blood pressure cu is properly wrapped around the arm according to the operang instrucons, if the cu