EasyManuals Logo

Gima PC-3000 User Manual

Gima PC-3000
Go to English
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #279 background imageLoading...
Page #279 background image
91
ENGLISH
NOTE: UT is the a.c. mains voltage prior to applicaon of the test level.
Table 3
Guidance and manufacturers declaraon – electromagnec immunity
The Paent Monitor is intended for use in the electromagnec environment specied below. The customer or the user
of The Paent Monitor should assure that it is used in such an electromagnec environment.
Immunity test IEC60601 test level Compliance level Electromagnec environment -guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
0,15MHz 80MHz
3 V RMS outside the ISM
band, 6 V RMS in the ISM
80 MHz to 2.7 GHz
3V/m
0,15MHz 80MHz
3 V RMS outside the ISM
band, 6 V RMS in the ISM
80 MHz to 2.7 GHz
3V/m
Portable and mobile RF communicaons
equipment should be used no closer
to any part of The Paent Monitor,
including cables, than the recommended
separaon distance calculated from the
equaon applicable to the frequency of
the transmier.
Recommended separaon distance
d=1.2
P
d=1.2
P
80MHz to 800MHz
d=2.3
P
800MHz to 2.5GHz
Where P is the maximum output
power rang of the transmier in
was (W) according to the transmier
manufacturer and d is the recommended
separaon distance in metres (m). b
Field strengths from xed RF
transmiers, as determined by an
electromagnec site survey ,a should be
less than the compliance level in each
frequency range .b
Interference may occur in the vicinity of
equipment marked with the following
symbol.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gima PC-3000 and is the answer not in the manual?

Gima PC-3000 Specifications

General IconGeneral
BrandGima
ModelPC-3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals