91
ENGLISH
NOTE: UT is the a.c. mains voltage prior to applicaon of the test level.
Table 3
Guidance and manufacturers declaraon – electromagnec immunity
The Paent Monitor is intended for use in the electromagnec environment specied below. The customer or the user
of The Paent Monitor should assure that it is used in such an electromagnec environment.
Immunity test IEC60601 test level Compliance level Electromagnec environment -guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
0,15MHz 80MHz
3 V RMS outside the ISM
band, 6 V RMS in the ISM
80 MHz to 2.7 GHz
3V/m
0,15MHz 80MHz
3 V RMS outside the ISM
band, 6 V RMS in the ISM
80 MHz to 2.7 GHz
3V/m
Portable and mobile RF communicaons
equipment should be used no closer
to any part of The Paent Monitor,
including cables, than the recommended
separaon distance calculated from the
equaon applicable to the frequency of
the transmier.
Recommended separaon distance
d=1.2
P
d=1.2
P
80MHz to 800MHz
d=2.3
P
800MHz to 2.5GHz
Where P is the maximum output
power rang of the transmier in
was (W) according to the transmier
manufacturer and d is the recommended
separaon distance in metres (m). b
Field strengths from xed RF
transmiers, as determined by an
electromagnec site survey ,a should be
less than the compliance level in each
frequency range .b
Interference may occur in the vicinity of
equipment marked with the following
symbol.