91
ENGLISH
NOTE: UT is the a.c. mains voltage prior to applicaon of the test level.
Table 3
Guidance and manufacturer’s declaraon – electromagnec immunity
The Paent Monitor is intended for use in the electromagnec environment specied below. The customer or the user
of The Paent Monitor should assure that it is used in such an electromagnec environment.
Immunity test IEC60601 test level Compliance level Electromagnec environment -guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
0,15MHz – 80MHz
3 V RMS outside the ISM
band, 6 V RMS in the ISM
80 MHz to 2.7 GHz
3V/m
0,15MHz – 80MHz
3 V RMS outside the ISM
band, 6 V RMS in the ISM
80 MHz to 2.7 GHz
3V/m
Portable and mobile RF communicaons
equipment should be used no closer
to any part of The Paent Monitor,
including cables, than the recommended
separaon distance calculated from the
equaon applicable to the frequency of
the transmier.
Recommended separaon distance
d=1.2
d=1.2
80MHz to 800MHz
d=2.3
800MHz to 2.5GHz
Where P is the maximum output
power rang of the transmier in
was (W) according to the transmier
manufacturer and d is the recommended
separaon distance in metres (m). b
Field strengths from xed RF
transmiers, as determined by an
electromagnec site survey ,a should be
less than the compliance level in each
frequency range .b
Interference may occur in the vicinity of
equipment marked with the following
symbol.