Unable to detect HR
ECG cable and leads are connected to monitor and paent well, but HR is
unable to be detected. It may be caused by poor integrity of ECG signals.
Unable to detect SpO
2
SpO
2
probe is connected to monitor and paent well, but SpO
2
is unable
to be detected. It may be caused by poor plethysmogram
signals.
Low baery Low baery voltage
ECG Lead O The ECG electrodes or cable fell o
SpO
2
Probe O SpO
2
probe fell o
B - Status/Error during Monitoring
“Cu error” cu is not wrapped correctly, or is not connected
“Air leak” Air moving part, tube or the cu leak air.
“Pressure error” Unstable cu pressure or tangled cu tubing
“Signal weak” Very weak signal because of the cu, or the paent has very weak pulse
“Over extent” The measurement range exceeds 255 mmHg (Pediatric paent over 135 mmHg)
“Over moon”
The repeated measurement due to moving, excessive noise during the stepping inaon and
measuring pressure and pulse, e.g. during paent shaking moon
“Signal overow” Blood pressure amplier overow due to excessive movement
“Leak in gas run” Leaking during the pneumac device tesng
“System error” Abnormal condion of CPU, such as register overow, divided by zero
“Adult”
The blood pressure measuring now is in adult mode. In this case, it is not allowed to
monitoring Pediatric or neonatal paent. Otherwise, there may be serious danger to the
Pediatric monitored.
“Pediatric” The blood pressure module is now worked in Pediatric measuring mode.
“PROBE OFF” SpO
2
probe fell o
“LEADS OFF” The ECG electrodes or cable fell o
“LEARNING” Learning arrhythmia for 15 seconds
“DEMO” The monitor is displaying the demo waveforms, which are generated by the monitor itself.