9.4 Message Reference
151
99
Starting a test
Syntax
(Command)
:STARt
Starts a test in the READY state.
Example :STARt
Error
The execution of this command in a state other than the READY state or "test
setting state" causes an execution error.
In the optional functions, if the interface command START is invalid, the execu-
tion of this command causes an execution error.
Note that an execution error occurs when double action is set to "1: ON" on the
optional function setting screen.
In the optional function setting, an execution error will occur if “output voltage
limit value < test voltage value” is set for the withstand-voltage test’s output volt-
age limit.
Note
To start a test using this command, set "START" for the interface command to
"Start command valid", which is an optional function setting.
In this command, set Momentary Out in the optional function setting to “0:OFF”.
At “1:ON”, the test forcibly stops immediately after test starts.
See "9.5 Interface Command "START"" (p. 186)
Forcibly ends a test and releases the hold state
Syntax
(Command)
:STOP
In the TEST state, forcibly ends a test. Furthermore, releases the
Hold function and returns to the READY state.
However, in the optional functions, when "FAIL Mode Function" is set
to ON, the Hold function cannot be disabled by this command.
See "6.5 Limiting the FAIL Hold Cancellation (FAIL mode)" (p. 90)
Example :STOP
Error
In the optional functions, when "FAIL Mode Function" is set to ON, an execution
error will occur if the instrument is in the FAIL Hold Mode.