9.4 Message Reference
153
99
Sets and Queries the type of voltage for withstand-voltage tests
Syntax
(Command)
:CONFigure:WITHstand:KIND <AC50 / AC60>
Sets the type of voltage for withstand-voltage tests in the READY
state.
(Query)
:CONFigure:WITHstand:KIND?
Queries the type of voltage for withstand-voltage tests in the READY
state.
(Response)
<AC50 / AC60>
AC50.....................50 Hz AC
AC60.....................60 Hz AC
Example
(Command)
:CONFigure:WITHstand:KIND AC50
Sets the test frequency at 50 Hz AC.
(Query)
:CONFigure:WITHstand:KIND?
(Response)
AC50
The test frequency is set at 50 Hz AC.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.