9.4 Message Reference
155
99
Sets and Queries the lower-limit test value for withstand-voltage tests
Syntax
(Command)
:CONFigure:WITHstand:CLOWer <0.1 to 19.9 (NRf)>
Sets the lower-limit test value (unit: mA) for withstand-voltage tests, in
the READY state.
(Query)
:CONFigure:WITHstand:CLOWer?
Queries the lower-limit test value (unit: mA) for withstand-voltage
tests.
<0.1 to 19.9 (NR2)>
Example
(Command)
:CONFigure:WITHstand:CLOWer 0.1
Sets the lower-limit test value for withstand-voltage tests to 0.1 mA.
(Query)
:CONFigure:WITHstand:CLOWer?
(Response)
0.1
The lower-limit test value is 0.1 mA.
Error
The execution of this command in a state other than the READY state or "test
setting state" causes an execution error.
When the lower limit of the withstand-voltage test is set to ON, an execution
error will occur when "upper limit test value lower limit test value" is set.
Sets and Queries the test time for withstand-voltage tests
Syntax
(Command)
:CONFigure:WITHstand:TIMer <0.3 to 999 (NRf)>
Sets the test time (unit: s) for withstand-voltage tests, in the READY
state.
(Query)
:CONFigure:WITHstand:TIMer?
Queries the test time (unit: s) for withstand-voltage tests.
(Response)
<0.3 to 999 (NR1/NR2)>
Example
(Command)
:CONFigure:WITHstand:TIMer 30.0
Sets the test time for withstand-voltage tests to 30.0 s.
(Query)
:CONFigure:WITHstand:TIMer?
(Response)
30.0
The test time is set to 30.0 s.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.