9.4 Message Reference
157
99
Sets and Queries the initial voltage for ramp-up for withstand-voltage tests
Syntax
(Command)
:CONFigure:WITHstand:VINItial <0.0 to 1.0 (NR2)>
Set the initial voltage for the withstand test voltage ramp-up as a pro-
portion of the test voltage.
(Query)
:CONFigure:WITHstand:VINItial?
Queries the initial voltage for ramp-up for withstand-voltage tests.
(Response)
<0.0 to 1.0 (NR2)>
Example
(Command)
:CONFigure:WITHstand:VINItial 0.5
Set the ramp-up initial voltage as 0.5 times that of the test voltage.
(Query)
:CONFigure:WITHstand:VINItial?
(Response)
0.5
Ramp-up initial voltage is test voltage x 0.5.
Error
• The execution of this command in a state other than the READY state or "test
setting state" causes an execution error.
• Execution error appears when value other than 0 to 1.0 is set.
Sets and Queries the contact check upper-limit voltage for withstand-voltage tests
Syntax
(Command)
:CONFigure:WITHstand:CNHI <0.20 to 5.00 (NRf)>
Sets the contact check upper-limit voltage for withstand-voltage tests
in the READY state.
(Query)
:CONFigure:WITHstand:CNHI?
Queries the contact check upper-limit voltage for withstand-voltage
tests.
(Response)
<0.20 to 5.00 (NR2)>
Example
(Command)
:CONFigure:WITHstand:CNHI 1.00
Sets the contact check upper-limit voltage for withstand-voltage tests at 1.00
kV.
(Query)
:CONFigure:WITHstand:CNHI?
(Response)
1.00
The contact check upper-limit voltage for withstand-voltage tests is set to 1.00
kV.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.