9.4 Message Reference
158
Sets and Queries the contact check lower-limit voltage for withstand-voltage tests
Syntax
(Command)
:CONFigure:WITHstand:CNLO <0.20 to 5.00 (NRf)>
Sets the contact check lower-limit voltage for withstand-voltage tests
in the READY state.
(Query)
:CONFigure:WITHstand:CNLO?
Queries the contact check lower-limit voltage for withstand-voltage
tests.
(Response)
<0.20 to 5.00 (NR2)>
Example
(Command)
:CONFigure:WITHstand:CNLO 1.00
Sets the contact check lower-limit voltage for withstand-voltage tests at 1.00
kV.
(Query)
:CONFigure:WITHstand:CNLO?
(Response)
1.00
The contact check lower-limit voltage for withstand-voltage tests is set to 1.00
kV.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.
Enables/disables and Queries the lower-limit test value for withstand-voltage tests
Syntax
(Command)
:WITHstand:CLOWer <ON/ OFF>
Enables and disables the lower-limit test value for withstand-voltage
tests, in the READY state.
(Query)
:WITHstand:CLOWer?
Queries the lower-limit test value enablement for withstand-voltage
tests.
(Response)
<ON/ OFF>
Example
(Command)
:WITHstand:CLOWer ON
Enables the lower-limit test value for a withstand-voltage test.
(Query)
:WITHstand:CLOWer?
(Response)
ON
The lower-limit test value for the withstand-voltage tests is set to ON.
Error
The execution of this command in a state other than the READY state or "test
setting state" causes an execution error.
When the upper limit and lower limit values of the withstand-voltage test are set
as "upper limit test value lower limit test value”, an execution error will occur
when the lower limit of the withstand-voltage test is set to ON.