9.4 Message Reference
160
Enables/disables and Queries the ramp-up time for withstand-voltage tests
Syntax
(Command)
:WITHstand:UTIMer <ON/ OFF>
Sets the ramp-up time ON/OFF for withstand-voltage tests in the
READY state.
(Query)
:WITHstand:UTIMer?
Queries the ramp-up time ON/OFF for withstand-voltage tests.
(Response)
<ON/ OFF>
Example
(Command)
:WITHstand:UTIMer ON
Sets the ramp-up time for the withstand-voltage tests to ON.
(Query)
:WITHstand:UTIMer?
(Response)
ON
The ramp-up time for the withstand-voltage tests is set to ON.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.
Enables/disables and Queries the ramp-down time for withstand-voltage tests
Syntax
(Command)
:WITHstand:DTIMer <ON/ OFF>
Sets the ramp-down time ON/OFF for withstand-voltage tests in the
READY state.
(Query)
:WITHstand:DTIMer?
Queries the ramp-down time ON/OFF for withstand-voltage tests.
(Response)
<ON/ OFF>
Example
(Command)
:WITHstand:DTIMer ON
Sets the ramp-down time for the withstand-voltage tests to ON.
(Query)
:WITHstand:DTIMer?
(Response)
ON
The ramp-down time for the withstand-voltage tests is set to ON.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.