9.4 Message Reference
161
99
Enables/disables and Queries the contact check upper-limit voltage for withstand-volt-
age tests
Syntax
(Command)
:WITHstand:CNHI <ON/ OFF>
Sets the contact check upper-limit voltage ON/OFF for withstand-volt-
age tests in the READY state.
(Query)
:WITHstand:CNHI?
Queries the contact check upper-limit voltage ON/OFF for withstand-
voltage tests.
(Response)
<ON/ OFF>
Example
(Command)
:WITHstand:CNHI ON
Sets the contact check upper-limit voltage for the withstand-voltage tests to
ON
.
(Query)
:WITHstand:CNHI?
(Response)
ON
The contact check upper-limit voltage for the withstand-voltage tests is set to
ON.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.
Enables/disables and Queries the contact check lower-limit voltage for withstand-volt-
age tests
Syntax
(Command)
:WITHstand:CNLO <ON/ OFF>
Sets the contact check lower-limit voltage ON/OFF for withstand-volt-
age tests in the READY state.
(Query)
:WITHstand:CNLO?
Queries the contact check lower-limit voltage ON/OFF for withstand-
voltage tests.
(Response)
<ON/ OFF>
Example
(Command)
:WITHstand:CNLO ON
Sets the contact check lower-limit voltage for the withstand-voltage tests to
ON
.
(Query)
:WITHstand:CNLO?
(Response)
ON
The contact check lower-limit voltage for the withstand-voltage tests is set to
ON.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.