9.4 Message Reference
174
Loads the set-value memory for insulation-resistance tests
Syntax
(Command)
:MEMory:INSulation:LOAD <1 to 8 (NR1)>
Loads the data saved for the insulation-resistance test file numbers
specified.
Example :MEMory:INSulation:LOAD 1
Loads memory file 1 for insulation-resistance tests.
Error
The execution of this command in a state other than the insulation-resistance
mode READY state causes an execution error.
The execution of this command in the withstand-voltage mode also causes an
execution error.
An execution error occurs when file numbers other than 1 to 8 are specified.
Saves set values for insulation-resistance tests in memory
Syntax
(Command)
:MEMory:INSulation:SAVE <1 to 8 (NR1)>
Saves the insulation-resistance test settings in the memory file num-
bers specified.
Example :MEMory:INSulation:SAVE 2
Saves the settings for insulation-resistance tests in memory file 2.
Error
The execution of this command in a state other than the insulation-resistance
mode READY state causes an execution error.
The execution of this command in the withstand-voltage mode also causes an
execution error.
An execution error occurs when file numbers other than 1 to 8 are specified.
Clears the set-value memory for insulation-resistance tests
Syntax
(Command)
:MEMory:INSulation:CLEar <1 to 8 (NR1)>
Clears and resets the data saved for the insulation-resistance test file
numbers specified.
Example :MEMory:INSulation:CLEar 3
Clears and resets the contents of memory file 3 for insulation-resistance tests.
Error
The execution of this command in a state other than the READY state causes
an execution error.
An execution error occurs when file numbers other than 1 to 8 are specified.