EasyManua.ls Logo

Keithley 2700 - :DATE Command for Calibration Date; :NDUE Command for Due Date

Keithley 2700
500 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-52 Basic DMM Operation Model 2700 Multimeter/Switch System User’s Manual
Reference
a. FUNCtion <name> [, <clist>]
Note that the <name> parameters in the table are enclosed in single quotes
(‘ ’). However, double quotes (“ ”) can instead be used. For example:
FUNC ‘VOLT:AC’ = FUNC “VOLT:AC”
Scan configuration — When using the <clist> command to configure a scan
channel, the scan channel must first be set to the appropriate function before
sending other commands to configure it. For example, to set scan channel 101
to use offset-compensated ohms, the following command sequence would be
sent:
FUNC 'FRES', (@101) ' Set scan channel 101 to Ω4 function.
FRES:OCOM ON, (@101) ' Enable offset-compensated ohms for scan
' channel 101.
If scan channel 101 was not set first for the Ω4 function, the following errors
will occur for the following operations:
Error -221 (Settings conflict) will occur when trying to enable offset-
compensated ohms for that channel.
Error +700 (Invalid function in scanlist) will occur when trying to query
the state of offset-compensated ohms for that channel:
FRES:OCOM ON? (@101)
Details on scanning are provided in Section 7.
b. FRESistance:OCOMpensated <b> [, <clist>]
The instrument does not have to be on the Ω4 function to enable offset-
compensated ohms. When Ω4 is selected, offset-compensated ohms will be
enabled.
When using the <clist> parameter to configure a scan channel for offset-
compensated ohms, that channel must already be set for Ω4. If set for another
function, a settings conflict (-221) will occur.
c. TEMPerature:TCouple:RJUNction:RSELect <name> [, <clist>]
To use the EXTERnal reference junction, the scan channel (101, 201, 301,
401, or 501) to be used for the measurement must already be configured to use
a thermistor or 4-wire RTD transducer. Otherwise, a settings conflict error
(-221) will occur. The TEMPerature:TRANsducer command is used to select
the transducer.
NOTE The following command can instead be used to select the reference
junction:
TEMPerature:RJUNction:RSELect <name> [, <clist>]

Table of Contents

Other manuals for Keithley 2700

Related product manuals